MIL-PRF-55365H
4.7.20 Weibull FRL grading (see 3.23). Capacitors shall be tested in accordance with
method 108 of MIL-STD-202. The following details and exceptions shall apply:
a. Distance of the heating source from specimens, in inches: Not applicable.
b. Method of mounting: Capacitors shall be mounted by their terminations.
c. Test temperature and tolerance: +85°C +4°C, -0°C.
d. Operating conditions: Accelerated dc voltage, +4 percent, -0 percent, as applicable (see table XII), shall be
applied gradually (not to exceed 5 minutes by a slow buildup of the voltage). Maximum nominal acceleration
factor (see table XII) shall be 20,000:1. Only the capacitors used in 4.7.20e must be fused. Voltage shall be
applied continuously, except for failure count periods. The impedance of the voltage source, as seen from
the terminals of each capacitor, shall not exceed 1 ohm. An electronic power supply capable of supplying at
least 5 amperes when a capacitor is shorted shall be used. A 1 ampere to 2 ampere fuse shall be
connected in series with each capacitor. Slow-blow fuses shall not be used. The manufacturer shall be able
to demonstrate that all monitored and unmonitored parts receive correct voltage. When separate equipment
is used for testing the monitored sample and the unmonitored sample (rest of the lot), the following is
required:
(1) Equipment shall be cross calibrated for temperature, voltage, and test duration to ensure equivalent
test conditions.
(2) The ramp rate of voltage application at the start of test shall be identical for monitored and
unmonitored test groups.
e. Minimum sample size for monitoring at the beginning of test prior to infant mortality period: 300 pieces, or
100 percent, whichever is less. If the sample size is less than 100 percent, the remainder shall be subject to
the same accelerated dc voltage for the same amount of time.
f. Duration of test: 40 hours minimum. The manufacturer shall record the test start and stop times for each lot
tested.
Timing:
x2
x3
Infant mortality period
x1
0 .................................0.......................... 0 ..........................0 ......................... 0
5-15
See
40 hours
See
4.7.20 i
minimum
4.7.20 i
minutes
g. Failure definition: A failure is defined as a blown fuse or equivalent.
h. Capacitors from positions with blown fuses: Capacitors from positions with blown fuses shall be removed
and shall not be delivered on the contract. These capacitors may be used for solderability testing or other
destructive testing.
i.
Failure count during test: The lot size (see 4.6.1.1.2) to be graded is established after removal of gross
defectives (infant mortality) (5 minutes - 15 minutes). The first failure count shall be performed at 2 hours
±0.3 hour after the test was started. If there are no failures at time x1, the manufacturer may use one of the
following options:
(1) Complete a minimum of 40 hours and compute the failure from MIL-STD-690, table II FRSP-90, based
on the number of failures at time x2, or;
(2) Extend time x1 from 2 hours to a maximum of 10 hours. A failure cannot be assumed. If there are still
no failures, option 4.7.20i(1) shall be used, or;
29
For Parts Inquires submit RFQ to Parts Hangar, Inc.
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business